| Model Number |
BLIS-X1 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
09/13/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not available for evaluation.The investigation is ongoing and a follow-up report will be submitted upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during the implantation of an intraocular lens (iol) into the right eye, the capsular bag collapsed when the lens was placed into the eye.The iol was removed, and a backup lens of a different model was successfully implanted.The patient outcome is good.
|
| |
|
Additional Manufacturer Narrative
|
|
Although requested, the device was not returned for evaluation.A lot number was not provided; therefore a device history record (dhr) review could not be completed.The trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the information provided, a root cause could not be determined.
|
| |
|