| Model Number |
SUTR ANCH HIP PEEK P-LCK 2.9X12.5MM |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 11/23/2023, it was reported by an arthrex subsidiary employee via (b)(4), that an ar-2923phs suture anchor peek hip pushlock anchor backed out.This occurred during a hip arthroscopy procedure on 11/08/2023 the material was safely removed and reassembled with the correct position and used to successfully complete the case.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely causes of the reported failure are improper bone preparation, misaligned insertion, and prying/leveraging the device during insertion.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h6.
|
| |
|
Event or Problem Description
|
|
Additional information was provided on 1/3/2024 where the arthrex subsidiary employee stated the guide was used accompanied by a drill bit to prepare the bone.The patient's bone quality was good.The implant was dismantled and all the remains of material that remained were recovered, without leaving anything inside the case.
|
| |
|