| Model Number |
GI100 |
| Medical Device Problem Codes |
Product Quality Problem (1506); Protective Measures Problem (3015)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); Laceration(s) (1946)
|
| Date of Event |
11/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A1-a6: not applicable.No patient involvement.This is an isolated incident as all current omnis instruments are manufactured with the guards installed.It was confirmed that there are only two old z-series omnis instruments in the field missing the guards, one of them was the instrument involved in this incident.Additionally, it has been confirmed that guards have been placed on both z-series instruments to prevent a recurrence of this incident.
|
| |
|
Event or Problem Description
|
|
Agilent field service engineer (fse) states that her left hand was grazed/scrapped by the rotating fan which was missing guards.This injury occurred between her thumb and index finger.There were 3-4 cuts (less than an inch long) and the bleed was stopped once a bandage was applied.The engineer confirmed that her injury has healed.Investigation confirmed that the instrument is a z-series production and was not manufactured with guards on the back side of the bulk fans.The instrument has been serviced and guards were placed on the fans.The instrument is fully operational and is available to the user.This is an internal agilent instrument, no patient diagnosis involved.
|
| |
|