EXACTECH, INC. EQUINOXE CAGE GLENOID MEDIUM, BETA; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
EQUINOXE CAGE GLENOID MEDIUM, BETA |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
04/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Section d10: concomitant products: equinoxe, humeral head expanded, 50mm (beta) (cat#: 310-03-50 / serial#: (b)(6).Equinox square torque define screw drive kit (cat#: 300-20-02 / serial#: (b)(6).Glnd kwire (cat#: 315-35-00 / serial#: (b)(6).Equinoxe replicator plate 4.5mm o/s (cat#: 300-10-45 / serial#: (b)(6).Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
As reported, 62 yo male patient primary anatomic surgery was on (b)(6) 2014.Patient was revised and converted to a reverse on (b)(6) 2022, in which the original stem was kept.The surgeon mentioned that the patient disappeared or went off for a while post first revision in 2022, alluding to maybe him doing something that caused the implant to break, but nothing specific is known.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: a, b, c, d, e, g.The reason for the revision reported cannot be confirmed from the information provided but may be due to inclusion of the polyethylene in the packaging recall, wear, loosening, infection, fracture, instability, dislocation, rotator cuff tear, and/or patient related factors.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|