| Model Number |
45-751019 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was stated to be available for return to the manufacturer for evaluation, but has not yet been returned.The alleged product issue/event as described could not be confirmed.If the device is received at a later date, an investigation will be performed and a supplemental mdr will be submitted.A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with the use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during retraction of the coil through the catheter the coil spontaneously detached.The procedure was completed.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation conclusion: the investigation of the returned coil system found the implant stretched at the proximal section, kinked, deformed, and separated from the pusher; however, no indications of activation using a detachment controller was found on the pusher heater coil.The investigation found the pusher's monofilament with a tensile break shape at the tip, which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the implant damage, but the damage is consistent with the device experiencing forces over specification.The catheter used in the procedure was not returned for evaluation so the investigation could not determine if it had caused or contributed to the reported complaint.
|
| |
|
Search Alerts/Recalls
|