• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE 6 MM CYLINDRICAL BUR; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE 6 MM CYLINDRICAL BUR; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB10036
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a cori assisted ukr, while removing the patient's bone using exposure mode with a real intelligence 6 mm cylindrical bur, the patient's bone suddenly would not be removed; the drill was checked and the bur was broken inside the drill.No pieces of the device fell into the patient.The procedure was resumed, without any delay, using a backup robotic drill.No patient injuries or complications were reported due to this issue.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(6).
 
Additional Manufacturer Narrative
H3, h6: the real intelligence 6 mm cylindrical bur, part number rob10036, serial number unknown, used for treatment was not returned for evaluation, therefore a device analysis was unable to be performed.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with device wear and tear due to repeated use.Based on the investigation, no containment or corrective action is recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Additional Manufacturer Narrative
H3, h6: an incomplete portion of a real intelligence 6 mm cylindrical bur, part # (b)(6), used for treatment, was returned for evaluation.As only part of the device was returned, a lot # could not be verified.The reported problem was visually confirmed.The bur fragment was found stuck inside of the collet of the returned drill (b)(6).The fracture was observed to occur at the smallest diameter location on the bur.Functional evaluation could not be performed due to broken/damaged parts.While the bur was confirmed to be broken, a definitive cause for the fracture could not be determined.Possible causes may have been related to a material defect, or due to a manufacturing fault.Additional factors which may have contributed to the failure include stresses applied to the bur, such as axial load applied from horizontal movements or compressive stress applied when plunging into bone.There seems to have been no fault to the returned drill.If the remaining portion of the bur is returned for investigation, the complaint can be reopened.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was not completed.The product was not returned and no evidence was made available to link the complaint to an escalation event.The failure mode and associated risk have been anticipated within the risk file including a documented risk level.Although no further containment or corrective action is recommended or required at this time, the failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE 6 MM CYLINDRICAL BUR
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key18370327
Report Number3010266064-2023-00239
Device Sequence Number20761372
Product Code OLO
UDI-Device Identifier00885556757512
UDI-Public885556757512
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberROB10036
Device Lot Number1125569
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2023
Supplement Date Received by Manufacturer01/05/2024
04/05/2024
Initial Report FDA Received Date12/20/2023
Supplement Report FDA Received Date01/08/2024
04/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
REAL INTELLIGENCE CORI, SN: (B)(6).; REAL INTELLIGENCE ROBOTIC DRILL, SN:(B)(6).; RI ROBOTIC DRILL ATTACHMENT, SN:(B)(6).
Patient SexFemale
-
-