It was reported that on november 29, 2023, a 14mm x 2.4 cortex screws would not fit into predrilled hole during a distal radius fixation case without substantial difficulty.Significant force was needed to get the screw to begin threading into bone.Surgeon and resident complained.Case was able to proceed successfully.Surgeon used another screw and impact to begin screw insertion.There was no surgical delay and no patient consequence.This report involves one 2.4mm cortex screw slf-tpng with t8 stardrive recess 14mm.This is report 1 of 1 for (b)(4).
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2b: additional device product codes: hrs.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|