| Catalog Number |
03666794001 |
| Medical Device Problem Code |
Off-Label Use (1494)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A display check was performed there were missing segments in the middle of the screen and at the top of the screen.There was a little corrosion found in the battery compartment around the coils.The meter was requested to be returned for investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of a display issue with the coaguchek xs meter.The reporter stated when the meter is powered on, she only sees segments at the top left and top right of the screen.When test strips are inserted, she can see segments in the middle of the screen and at the top left and top right of the screen.However, some of the segments were flashing.The reporter alleged she could not read any numbers as she only sees segments that weare part of the numbers.There was no allegation of an actual result misinterpretation.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect meter was received and disassembled for further investigation.It was found the circuit board was contaminated and corroded by penetrated liquid.This contamination also affected the conductive rubber contacts the root cause of the display issue was contamination of the contacts due to improper user handling or maintenance.There was no malfunction of the device.
|
| |
|