• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA CUP: VERSAFITCUP CC TRIO ACETABULAR SHELL CC TRIO Ø 48; HIP CEMENTLESS ACETABULAR SHELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA CUP: VERSAFITCUP CC TRIO ACETABULAR SHELL CC TRIO Ø 48; HIP CEMENTLESS ACETABULAR SHELL Back to Search Results
Catalog Number 01.26.45.0048
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 11/27/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 04 december 2023 lot 1910233: (b)(4) items manufactured and released on 31-march-2020.Expiration date: 2025-03-16.No anomalies found related to the problem.To date, all items of the same lot have been sold without any similar reported event during the period of review.Clinical evaluation performed by medical affairs department: revision of the cup performed at about 2 years 2 months after the primary, due to medialization of the acetabular cup and suboptimal offset.According to report, there was no loosening nor infection.From the radiographic image, the cup seems slightly medialized but the medial wall is intact.We cannot tell if this is the result of post-surgery migration or the surgeon's choice: in the latter case, no explanation for this decision was supplied.However, without a controlateral image it is difficult to evaluate the position of the cup.There is no reason to suspect a faulty device.
 
Event or Problem Description
At about 2 years 2 months after the primary, revision surgery performed on the left side due to medialization of the acetabular cup and suboptimal offset.Patient had discomfort since the primary surgery.No loosening of the implants.No infection.The surgeon revised successfully the cup, liner and head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUP: VERSAFITCUP CC TRIO ACETABULAR SHELL CC TRIO Ø 48
Common Device Name
HIP CEMENTLESS ACETABULAR SHELL
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key18372060
Report Number3005180920-2023-01020
Device Sequence Number7710782
Product Code LZO
UDI-Device Identifier07630030807756
UDI-Public07630030807756
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K103352
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/16/2025
Device Catalogue Number01.26.45.0048
Device Lot Number1910233
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2023
Initial Report FDA Received Date12/21/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/31/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexFemale
Patient Weight66 KG
-
-