• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLSTENT ENTERAL ENDOPROSTHESIS; STENT, COLONIC, METALIC, EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLSTENT ENTERAL ENDOPROSTHESIS; STENT, COLONIC, METALIC, EXPANDABLE Back to Search Results
Model Number M00565590
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 11/29/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to boston scientific corporation that a wallstent enteral uncovered stent was to be implanted in the intestinal tract to treat a malignant 4 cm intestinal obstruction during a stent implantation procedure performed on (b)(6) 2023.The patient's anatomy was moderately tortuous.During the procedure, the burrs at the tip of the stent caused intestinal perforation.The patient was sent for surgery to address the perforation and remove the stent.The procedure was completed with another wallstent enteral uncovered stent.The patient's condition after the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: imdrf patient code e2114 captures the reportable patient complication of perforation.Imdrf impact code f19 captures the surgical procedure performed to address the perforation.
 
Additional Manufacturer Narrative
Block h6: imdrf patient code e2114 captures the reportable patient complication of perforation.Imdrf impact code f19 captures the surgical procedure performed to address the perforation.Block h10: the wallstent enteral uncovered stent delivery system was returned for analysis.The stent was not received for analysis.Visual inspection found that the outer sheath and the inner sheath were kinked.No other damages were noted with the delivery system.Product analysis confirmed the reported events of perforation and surgical intervention as known potential adverse events associated with the use of the device and are documented in the labeling (including both short- or long-term known complications or adverse reactions).Additionally, the additional investigation findings of outer sheath and inner sheath kinking were most likely due to procedural factors such as lesion characteristics, the handling of the device, and the technique used by the physician (force applied), which could have resulted in the damages noted to the device.Therefore, a review and analysis of all available information indicated that the most probable cause is known inherent risk of device.A product labeling review identified that the device was used per the instructions for use (ifu)/product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a wallstent enteral uncovered stent was to be implanted in the intestinal tract to treat a malignant 4 cm intestinal obstruction during a stent implantation procedure performed on (b)(6) 2023.The patient's anatomy was moderately tortuous.During the procedure, the burrs at the tip of the stent caused intestinal perforation.The patient was sent for surgery to address the perforation and remove the stent.The procedure was completed with another wallstent enteral uncovered stent.The patient's condition after the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT ENTERAL ENDOPROSTHESIS
Common Device Name
STENT, COLONIC, METALIC, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18372071
Report Number3005099803-2023-06796
Device Sequence Number12657278
Product Code MQR
UDI-Device Identifier08714729285083
UDI-Public08714729285083
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K000281
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/10/2023
Device Model NumberM00565590
Device Catalogue Number6559
Device Lot Number0028533718
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/30/2023
Supplement Date Received by Manufacturer01/24/2024
Initial Report FDA Received Date12/21/2023
Supplement Report FDA Received Date02/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age59 YR
Patient SexMale
Patient Weight75 KG
-
-