| Model Number |
160002 |
| Medical Device Problem Code |
Misassembly by Users (3133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complaint has been reopened and reviewed according to fda form 483 inspectional observation ems #2, eobs2 from the fda inspection conducted between july 17 to july 21, 2022 at the ems and bonaduz sites.A detailed investigation was performed by an expert from the technical service: since the complaint in question was submitted to hamilton medical ag over 1 year ago, no attempt will be made to obtain additional information.No further investigation or correction will be performed except those mentioned above.In future hamilton medical ag will report an event similar to this issue as it will be deemed a reportable event.The allegation in this complaint was confirmed to be a complaint.With this investigation it has been confirmed that the device failed to meet its specifications at the time of the event while the ventilator was being prepared for use.The root cause of the problem was caused by the reverse connection of the blue and white monitoring tubes connected to the flow sensor.There was no patient or user harm.
|
| |
|
Event or Problem Description
|
|
The report from hospital say: on apr.11, 2022, in the 107 ward area, building 1, the medical staff turned on the ventilator but found that the machine could not work.After startup, the flow sensor showed an alarm "flow sensor reverse".Hamilton-c2 made in mar.9, 2018.
|
| |
|
Event or Problem Description
|
|
The report from hospital say: on (b)(6) 2022, in the 107 ward area, building 1, the medical staff turned on the ventilator but found that the machine could not work.After startup, the flow sensor showed an alarm "flow sensor reverse".Hamilton-c2 made in mar.9, 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint has been reopened and reviewed according to fda form 483 inspectional observation ems #2, eobs2 from the fda inspection conducted between july 17 to july 21, 2022 at the ems and bonaduz sites.A detailed investigation was performed by an expert from the technical service: since the complaint in question was submitted to hamilton medical ag over 1 year ago, no attempt will be made to obtain additional information.No further investigation or correction will be performed except those mentioned above.In future hamilton medical ag will report an event similar to this issue as it will be deemed a reportable event.The allegation in this complaint was confirmed to be a complaint.With this investigation it has been confirmed that the device failed to meet its specifications at the time of the event while the ventilator was being prepared for use.The root cause of the problem was caused by the reverse connection of the blue and white monitoring tubes connected to the flow sensor.There was no patient or user harm.Hamilton medical ag complaint number: (b)(4).Follow-up 1 - corrected information: udi related data quality updates only.Field d4 was updated with full udi information as requested.Updated fields: b4, d1, d4, g3, g6, h2, h4.
|
| |
|