• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit is showing electrical code #50.There was no harm reported.
 
Manufacturer Narrative
Corrected data: b5, b6, b7, d10, g2, h6 (clinical and impact code).Updated data: b4, e1 (initial reporter and phone#), e2, e3, g3, g6, h2, h10, h11.
 
Event Description
It was reported that during a routine check, the cs300 intra-aortic balloon pump (iabp) unit is showing electrical code #50.There was no patient involvement.
 
Manufacturer Narrative
Additional information: tel -(b)(6) & e1(event site postal code - (b)(6)) it was reported that the cs300 intra-aortic balloon pump (iabp) showing electrical code #50.There was no patient involvement and no patient harm reported.A getinge field service engineer fse was dispatched to the site to evaluate the unit.Checked and found that machine 5k hours maintanance kit is defective and it needs to be replaced.Quote was submitted but customer is not willing to repair it as per their distributor.H3 other text : distributor performed repair.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18372718
MDR Text Key331105541
Report Number2249723-2023-05414
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 12/11/2023
Initial Date FDA Received12/21/2023
Supplement Dates Manufacturer Received01/26/2024
05/02/2024
Supplement Dates FDA Received01/30/2024
05/02/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/19/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.; UNKNOWN.
-
-