| Model Number |
N/A |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Metal Related Pathology (4530)
|
| Date of Event |
12/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a explore radial head implant disassociated from stem after 3 years of being implanted.The surgeon revised, taking the dissociated head out which had the screw in it.The stem was stable and stayed in.He cleaned the female threads checked to make sure the threads weren¿t striped by using a screw and torquing on it.Nothing was cross threaded or stripped.He then retrialed and implanted a new head with a new screw.Attempts have been made and there is no further information at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10 - medical product: catalog #: 11-210063, explor 8x28mm impl stem w/scr, lot # 037350.H3: product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Proposed annex g code: mechanical (g04) - head.The reported event is confirmed by medical records/radiographs which were reviewed and identified the following: a radial head replacement implant is present.The articular portion is displaced from the stem and no longer articulates with the capitellum.There is no fracture.Visual examination of the returned product/provided pictures identified the explor implant head shows wear on bottom and has worn part of the etching.The articulating surface shows very little wear.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|