• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 6,4/7,8 FR. X 430 MM, 7°, 4,2 FR. CHANNEL, WITH WA00395A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH URETEROSCOPE, 6,4/7,8 FR. X 430 MM, 7°, 4,2 FR. CHANNEL, WITH WA00395A Back to Search Results
Model Number WA29040A
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus, the ureteroscope, 6,4/7,8 fr.X 430 mm, 7°, 4,2 fr.Channel, with wa00395a had a crack in the camera head connection.The issue was found during reprocessing for a therapeutic procedure that was completed using a similar device.There were no reports of patient injury or medical intervention associated with this event.
 
Additional Manufacturer Narrative
E1: full establishment name: (b)(6) hospital.The device was returned and evaluated, and the customer¿s allegation was confirmed, the eyepiece was damaged.Additional findings include the following: the outer tube was damaged; also, there were deposits on the inside of the eyepiece cover glass (stain).The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to wear and tear as well as excessive force.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETEROSCOPE, 6,4/7,8 FR. X 430 MM, 7°, 4,2 FR. CHANNEL, WITH WA00395A
Common Device Name
URETEROSCOPE, 6,4/7,8 FR. X 430 MM, 7°, 4,2 FR. CHANNEL, WITH WA00395A
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18372907
Report Number9610773-2023-03722
Device Sequence Number7579980
Product Code FGB
UDI-Device Identifier04042761062497
UDI-Public04042761062497
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K951855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA29040A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2023
Supplement Date Received by Manufacturer03/12/2024
Initial Report FDA Received Date12/21/2023
Supplement Report FDA Received Date03/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-