| Model Number |
KNOTLESS T-ROPE SYN-DESMOSIS REPR KIT,SS |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a syndesmosis surgery the knot on the enbo-buttons doesn't slide well at all.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to a manufacturing issue; however, due to the device not being returned, we are unable to confirm the reported event.
|
| |
|
Search Alerts/Recalls
|