| Catalog Number |
04906969001 |
| Medical Device Problem Code |
Computer Operating System Problem (2898)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
There was an allegation of regarding an alleged sample mismatch on a cobas u 411 urine analyzer.Reportedly, the sample results were assigned to the wrong patient.The customer stated that a sample was scanned and not measured immediately.A second sample was then scanned and the result of the second sample was assigned to the first sample.No incorrect patient results were reported outside of the laboratory.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
No product problem was identified.It was determined that handling and proper knowledge of the workflow configuration by the operator is essential to avoid mismatch.
|
| |
|