• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE GH; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE GH; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a tasp was returned fractured.Attempts to obtain additional information have been made; however, no more information is available.
 
Additional Manufacturer Narrative
(b)(4).H3- the product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual inspection of device found it to show wear and is fractured.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations/anomalies identified.A definitive root cause cannot be determined.However, the ztr wa 0820 20 documents the summary of investigations for complaints associated with the tibial articular surface provisional (tasp) fracture.Fractured tasps were analyzed by optical microscopy revealing hackle marks and river lines in the case of bending overload; and hackle marks, river lines and striations were found in the case of low cycle fatigue culminating in a bending overload failure type.This complaint is confirmed per device evaluation.If any further information is received which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.H6: component code: proposed component (annex g) code is: - mechanical (g04) - provisional bottom.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE GH
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18375589
Report Number0001822565-2023-03687
Device Sequence Number7579589
Product Code JWH
UDI-Device Identifier00889024245044
UDI-Public(01)00889024245044(10)63683147
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42527000707
Device Lot Number63683147
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/06/2023
Supplement Date Received by Manufacturer02/09/2024
Initial Report FDA Received Date12/21/2023
Supplement Report FDA Received Date02/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-