| Model Number |
MI1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.Ct imaging has been scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: based on the received information, a damage to the active electrode due to reported mechanical impact appears very likely.However, to determine an exact root cause device investigation would be necessary.No information on possible further steps has been received.
|
| |
|
Event or Problem Description
|
|
The user's hearing performance with the device is affected.Ct imaging has been scheduled.The user will be monitored.
|
| |
|
Search Alerts/Recalls
|