• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® POLYURETHANE URETERAL CATHETERS, OPEN TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® POLYURETHANE URETERAL CATHETERS, OPEN TIP Back to Search Results
Catalog Number 135005
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 11/22/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that during cystoscopy procedure, doctor noticed a defective area of three of the red, right sided, 5fr x69cm open tip urethral catheters.All from the same lot number.They explained that the end of the catheter with the screw cap attachment was crimped and could not attach guide wire to pass through.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned, and further investigation was not conclusive.Although a specific root cause cannot be determined, potential root cause for this event could be, ¿kinked catheter¿.The device history record review could not be performed without a lot number.The instructions for use were found adequate and state the following: "indications for use: the bard® polyurethane ureteral catheters are intended for use in the drainage of urine from the ureter.The polyurethane ureteral catheter is inserted through the urethra, into the bladder and up through the ureter.These catheters also allow the delivery of contrast media.Directions for use: ureteral catheters are designed in a variety of distal tip configurations.The anatomical characteristics of the ureteral orifice and physician preference should determine the choice of design in a specific clinical situation.Closed tip placement (placed without a guidewire) after placement of the cystoscope into the bladder, the ureteral orifice is located and examined and a choice of ureteral catheter is made.The catheter adapter is secured to the ureteral catheter, and the catheter is flushed to evacuate all air bubbles.Under direct vision, the catheter tip is positioned within the ureteral orifice.Contrast media is then injected through the ureteral catheter into the ureter.Under fluoroscopy, the ureter is visualized and required intervention determined.Open tip placement (placed with a guidewire) after placement of the cystoscope into the bladder, the ureteral orifice is located and examined and a choice of ureteral catheter is made.Under direct vision, the guidewire is positioned within the ureteral orifice.The ureteral catheter is inserted over the guidewire and advanced into the ureter.The guidewire is removed and the catheter adapter is secured to the catheter.Gentle aspiration to evacuate air from the catheter can be performed at this time.Contrast media is then injected through the ureteral catheter into the ureter.Under fluoroscopy, the ureter is visualized and required intervention determined.Warning: this unit should not be used as a ureteral splint.Warning: this is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Caution: do not withdraw ureteral catheter while it is deflected in endoscope.Avoid sharp bending.Caution: do not over-tighten the catheter adapter.Over-tightening the catheter adapter may occlude the lumen of the catheter.Caution: when using the tigertail¿ catheter without the stabilizing support of a guidewire, be aware that a malfunction may occur where the flexible tip may detach if excessive force is made in contact with the walls of the bladder, ureter or renal pelvis.Should the flexible portion of the catheter become detached, retrieve with an endourology grasping device." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during cystoscopy procedure, doctor noticed a defective area of three of the red, right sided, 5fr x69cm open tip urethral catheters.All from the same lot number.They explained that the end of the catheter with the screw cap attachment was crimped and could not attach guide wire to pass through.Also stated there was delay of procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® POLYURETHANE URETERAL CATHETERS, OPEN TIP
Common Device Name
URETERAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18376964
Report Number1018233-2023-09092
Device Sequence Number15405195
Product Code EYB
UDI-Device Identifier10801741126946
UDI-Public(01)10801741126946
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number135005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2023
Supplement Date Received by Manufacturer03/19/2024
Initial Report FDA Received Date12/21/2023
Supplement Report FDA Received Date03/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-