• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC INFINITY ADAPTIS TALDOME SZ3 INFINITY ADAPTIS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC INFINITY ADAPTIS TALDOME SZ3 INFINITY ADAPTIS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 33680023
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 09/27/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event could be confirmed, since evaluation of provided ct scans in collaboration with information obtained from the field find the device has been explanted.Microbiologist reviewed the sterilization procedures, environmental monitoring, bioburden data and the dhr and noted: the subject device was packaged according to established design and process specifications, and was sterilized according to procedure.No deviation for a non-conformance could be found.Upon further investigation of the ct scans by healthcare professionals the following was observed, ¿there is a girdlestone situation visible in the ct scan.Together with the information provided by the site, it is clear that there is an infection present, here.After more then 6 months it is not likely, that the infection came with the implantation or through the implant.Therefore, a patient related cause for the infection can be presumed." infections are often caused by a multitude of factors such as comorbidity of the patient, the trauma, the surgical intervention and the aftercare.While the investigation was able to confirm an infection was present, more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device disposition unknown.
 
Event or Problem Description
The digital stryker prophecy team received a ct scan indicating that the patient underwent a device explant procedure due to infection.The patient was implanted with a cement spacer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY ADAPTIS TALDOME SZ3 INFINITY ADAPTIS
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18378116
Report Number3010667733-2023-00802
Device Sequence Number7711051
Product Code HSN
UDI-Device Identifier00889797101547
UDI-Public00889797101547
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K172633
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number33680023
Device Lot Number1733370
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
-
-