| Catalog Number |
690-00-22E |
| Medical Device Problem Code |
Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Insufficient Information (4580)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported clicking in hip joint on movement after revision of left tr to constrained liner for recurrent dislocation.
|
| |
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding audible noise involving a trident liner was reported.The event was confirmed via provided video of the issue.Method & results: product evaluation and results: the device was not returned; however, an intraoperative video was provided.The video shows the bipolar portion of the constrained liner making a "popping" noise when the hip is articulated.The cause of the noise cannot be determined from the video.Clinician review: no medical records were received for review with a clinical consultant.Product history review: could not be performed as the device lot details were not provided.Complaint history review: could not be performed as the device lot details were not provided.Conclusions: it was reported that the patient was revised due to a "clicking" noise which was a nuisance to the patient.The device was not returned; however, an intraoperative video was provided.The video shows the bipolar portion of the constrained liner making a "popping" noise when the hip is articulated.The cause of the noise cannot be determined from the video.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
Patient reported clicking in hip joint on movement after revision of left tr to constrained liner for recurrent dislocation.
|
| |
|