AOMORI OLYMPUS CO., LTD. SINGLE USE ELECTROSURGICAL KNIFE KD-625 Back to Search Results
Model Number KD-625LR
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Perforation (2001); Stenosis (2263)
Date of Event 11/20/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Olympus reviewed the following literature titled "endoscopic dilation for stenosis after esophageal endoscopic submucosal dissection: knack and pitfalls".When a post-resection ulcer is expected to involve more than three-fourths of the esophageal circumference, it is advisable to consider preventive measures against stenosis, such as submucosal triamcinolone injection, as such lesions are associated with a high risk of stenosis following endoscopic resection.In the case of stricture development, endoscopic balloon dilation (ebd) is performed.To prevent perforation during ebd, it is recommended to use a 12-15 mm balloon and gradually increase the pressure to 8 atm over 4 min.For severe stenosis, a 10-12mm balloon is used.In the event of a perforation during ebd, fasting and administrating antibiotics are necesssary.If conservative treatment fails to show any improvement, surgical drainage should be considered.Type of adverse events/number of patients postoperative stenosis- 3 patients, perforation- 1 patient.This literature article requires 2 reports.The related patient identifiers are as follows: (b)(6) : kd-625lr.(b)(6) : gif-h290t.There is no report of any olympus device malfunction in any procedure described in this study.
 
Additional Manufacturer Narrative
Since the literature described "hook knife j" we selected ¿kd-625lr¿ as a representative product.The suspect device has not been returned to olympus for evaluation.The investigation is in process.*the literature article is attached for additional information.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the serial and/or lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the relationship between the device and the adverse event cannot be confirmed.There was no complaint reported on the subject device.There is no evidence of an olympus device malfunction.Therefore, the root cause cannot be determined.This supplemental report includes a correction to g2 to provide information that was inadvertently not included in the initial medwatch.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ELECTROSURGICAL KNIFE KD-625
Common Device Name
SINGLE USE ELECTROSURGICAL KNIFE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18379429
Report Number9614641-2023-01984
Device Sequence Number7712758
Product Code KNS
UDI-Device Identifier04953170407826
UDI-Public04953170407826
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K182408
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Study,Literature,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberKD-625LR
Device Lot NumberUNKNOWN(LITERATURE)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2023
Supplement Date Received by Manufacturer01/19/2024
Initial Report FDA Received Date12/21/2023
Supplement Report FDA Received Date01/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-