| Model Number |
36705 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the needle was deformed.The target lesion was located in the right inferior lung.A 3.0/17/15 leveen superslim was selected for use.During the procedure, after local anesthesia, the device was found to be deformed and curved when taken out of the patient.A new ablation needle was used to complete the procedure.There were no patient complications, and the patient was stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter address 1: (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that the needle was deformed.The target lesion was located in the right inferior lung.A 3.0/17/15 leveen superslim was selected for use.During the procedure, after local anesthesia, the device was found to be deformed and curved when taken out of the patient.A new ablation needle was used to complete the procedure.There were no patient complications, and the patient was stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter address 1: (b)(6).Device evaluated by mfr: the electrode device was returned with the rfa electrode cable; nevertheless, media was provided, and it was observed that the device was bent.Visual inspection revealed no damages at the handle device and it was observed that the needle was bent.Functional inspection was performed by extending and retracting the handle, and the needle/tines were activated without any problems.
|
| |
|