| Catalog Number |
01.26.45.0054 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
12/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision surgery due to overhang versafitcup cc trio cup.Cup, head and liner successfully revised.Primary info not available.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 20 december 2023: lot 2009363: (b)(4) items manufactured and released on 21-dec-2020.Expiration date: 2025-12-10.No anomalies found related to the problem.To date, all items of the same lot have been sold without any similar reported event during the period of review.Investigation performed by r&d knee manager: from the received images, it is not possible to determine the root cause of the event; the cup seems to be without any particular defect or damage, while the ceramic ball head was without any particular sign, except for the several scratches most probably related to the revision/removal.A possible reason can be related to the size of the implant, as the second chosen size revealed to be well fixed.
|
| |
|
Additional Manufacturer Narrative
|
|
Clinical evaluation performed on the (b)(6) 2024 by medacta clinical affairs manager: acetabular cup revised due to iliopsoas irritation.Most probably, this may have been caused by anterior overhang of the cup.From the radiographic images, the cup appears in suboptimal position.There is no reason to suspect a malfunctioning device.
|
| |
|