| Catalog Number |
129401060 |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Migration (4003)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Joint Laxity (4526)
|
| Date of Event |
12/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Revision left knee.Primary implants - right lcs femur, 12.5mm insert and size 5 tibia.Revision for subsidence and loosening of components.Femur removed by hand - clean (cement left on bone - no cement details available).Tibia removed very easily as well with cement all attached to bone and not implant dark metallosis appeared on knee tissue, no signs of infection.Surgeon washed, debrided then revised using attune revision.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : according to the information received: revision left knee.Primary surgery completed by [surgeon]- unknown surgery date and hospital.Primary implants - right lcs femur, 12.5mm insert and size 5 tibia.Revision for subsidence and loosening of components.Femur removed by hand - clean (cement left on bone - no cement details available).Tibia removed very easily as well with cement all attached to bone and not implant.Dark metallosis appeared on knee tissue, no signs of infection.Surgeon washed, debrided then revised using attune revision.The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed no cement was attached to lcs complete fem cem r lg, and signs of burnishing were observed on device too.Therefore, we were able to confirm loosening and micromotion on device.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint is confirmed as the observed condition of the lcs complete fem cem r lg would contribute to the complained device issue.Based on the investigation findings it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : a manufacturing record evaluation was performed for the finished device 129401060/3583834 number, and no nonconformances/manufacturing irregularities were identified.Device history review : a manufacturing record evaluation was performed for the finished device 129401060/3583834 number, and no nonconformances/manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|