SMITH & NEPHEW, INC. SPEED PIN QK CONNECT ADAPTER; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
74013489 |
| Medical Device Problem Codes |
Fracture (1260); Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a tka surgery the speed pin qk connect adapter itself broke into the gii mis dcf distal cut blk, so it also became unusable.The procedure had been completed, without any delay, using the same device.No injury was reported as a consequence of this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the devices were not returned for evaluation; therefore, a device analysis could not be performed.Device batch numbers were not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history for the previous 12 months did not reveal similar events for the listed devices.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.At this time, we have no reason to suspect that the products failed to meet any specifications at the time of manufacture.These devices are reusable instruments that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|