| Model Number |
39202 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that difficulty in removing from guidewire occurred.The 100% stenosed target lesion was located in severely tortuous and severely calcified common iliac artery.After crossing the lesion with a non-bsc guidewire, a 9 x 100 x 75 epic stent was advanced and successfully deployed.An attempt to withdraw the delivery system was made, but it was difficult to remove and was stuck together with the non-bsc guidewire.Both devices were removed and the procedure was completed with another of the same device.There were no patient complications nor injuries reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - device codes: corrected to 'entrapment of device' and 'difficult to remove.
|
| |
|
Event or Problem Description
|
|
It was reported that difficulty in removing from guidewire occurred.The 100% stenosed target lesion was located in severely tortuous and severely calcified common iliac artery.After crossing the lesion with a non-bsc guidewire, a 9 x 100 x 75 epic stent was advanced and successfully deployed.An attempt to withdraw the delivery system was made, but it was difficult to remove and was stuck together with the non-bsc guidewire.Both devices were removed and the procedure was completed with another of the same device.There were no patient complications nor injuries reported.
|
| |
|