MAUDE Adverse Event Report: INTEGRA LIFESCIENCES MANSFIELD CERELINK ICP MONITOR; ICP MONITOR-CERELINK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES MANSFIELD CERELINK ICP MONITOR; ICP MONITOR-CERELINK Back to Search Results
Catalog Number 826820
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported a cerelink monitor ((b)(6).) showing out of range values and continued to occur after changing the cables, power supply and moving it away from other devices.Over the weekend, the value became negative (-4).Troubleshooting was done by electrocardiogram (ecg) lead changed, machine plugged into wall and keeping it away from other devices however, it was not resolved.Eventually the numbers became positive but wide swings in the values 2-30, with the low values not in line with the imaging and clinical activities.No patient injury reported.Customer stated that it may have been user error.No additional information available.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h6, h11.The cerelink monitor (id (b)(4) was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined; however, the customer reported that they did not feel a need to continue pursuing an investigation as they believe there may have been user error causing the issue.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Updated fields: b5, d8, d9, g1, g3, g6, h1, h2, h3, h9.Follow-up submitted to include "correction/removal rpt number" (3014334038-10/05/2023-002-r) (h9).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERELINK ICP MONITOR
Common Device Name
ICP MONITOR-CERELINK
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 08540
6099362319
MDR Report Key18383947
Report Number3014334038-2023-00195
Device Sequence Number14505644
Product Code GWM
UDI-Device Identifier10886704075622
UDI-Public10886704075622
Combination Product (Y/N)N
PMA/510(K) Number
K183406
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number826820
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/05/2023
Supplement Date Received by Manufacturer05/28/2024
07/11/2024
Initial Report FDA Received Date12/22/2023
Supplement Report FDA Received Date06/16/2024
07/12/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3014334038-10/05/2023-00
Patient Sequence Number1
Patient SexUnknown
-
-