| Model Number |
3043 |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1 : (b)(6).H3: device not received by manufacturer.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the oximetry probe's sensors stopped working within a short period of time.The spo2 values were inconsistent with the patient's condition.The event occurred during patient use and there was no patient harm/adverse event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device not returned to date.H6 - evaluation codes: updated.Device evaluation: no product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.The exact cause could not be determined; however, based on the reported problem, the most probable cause could be improper patient attachment tape use.A device history report (dhr) review could not be performed as the dhr is at the supplier.If the product is returned, the manufacturer will reopen this complaint for further investigation.H3 - device returned to manufacturer: no.
|
| |
|
Additional Manufacturer Narrative
|
|
D4: updated **udi related data quality updates only**.
|
| |
|