| Model Number |
101/860/085 |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that air escaped from the cuff.When the cuff was before using it, the cuff deflated.During the patient's intubation, the cuff seemed to be deflated, making it easier for the tube to fall out.There was no patient involvement and no patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: other; device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
D9 - date returned to mfg; 1/2/2024.H3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One used decontaminated device sample was received for investigation without its original packaging.Under visual inspection the sample appeared to be in good condition.An inflation test was performed on the received sample.It was confirmed that after twelve (12) hours the cuff was still fully inflated.The complaint was not confirmed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.No action taken and no trend of similar customer complaints was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
D4 was corrected.**udi related data quality updates only**.
|
| |
|