MAUDE Adverse Event Report: SMITHS MEDICAL CZECH REPUBLIC A. S PORTEX TUBES BLUSELECT SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL CZECH REPUBLIC A. S PORTEX TUBES BLUSELECT SUCTIONAID; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/860/085
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/24/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that air escaped from the cuff.When the cuff was before using it, the cuff deflated.During the patient's intubation, the cuff seemed to be deflated, making it easier for the tube to fall out.There was no patient involvement and no patient harm/adverse event reported.
 
Additional Manufacturer Narrative
H3: other; device not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Additional Manufacturer Narrative
D9 - date returned to mfg; 1/2/2024.H3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One used decontaminated device sample was received for investigation without its original packaging.Under visual inspection the sample appeared to be in good condition.An inflation test was performed on the received sample.It was confirmed that after twelve (12) hours the cuff was still fully inflated.The complaint was not confirmed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.No action taken and no trend of similar customer complaints was identified.
 
Additional Manufacturer Narrative
D4 was corrected.**udi related data quality updates only**.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUSELECT SUCTIONAID
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL CZECH REPUBLIC A. S
olomoucká 306
hranice 753 0 1
EZ  753 01
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key18384542
Report Number3011237704-2023-00014
Device Sequence Number19522964
Product Code BTO
UDI-Device Identifier15019517076165
UDI-Public(01)15019517076165(17)280223(10)4310532
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K173384
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number101/860/085
Device Catalogue Number101/860/085CZ
Device Lot Number4310532
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2023
Supplement Date Received by Manufacturer01/24/2024
01/24/2024
Initial Report FDA Received Date12/22/2023
Supplement Report FDA Received Date02/20/2024
07/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-