| Catalog Number |
223600002 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Deformation (2976); Naturally Worn (2988); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).B3: date of event is an unknown date.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the plastic guides are damaged and will need replacing.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained, that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received.A.Can you please clarify, what do you mean by ¿damaged¿? was it worn, cracked, broken into pieces, bent, stripped, cross threaded or any device interaction.Worn away.The hex components of the guide have been burred out, so that they no longer fit onto the hex component of the guide handle.B.Was there a surgical delay? if yes, what is the duration of the delay? no.C.Can you confirm, the event date once? unsure.Gradual wearing over time, so no specific date, only informed date.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = according to the information received, "informed via email that the plastic guides are damaged and will need replacing" the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that the device 223600002, apg+ sizer pin guide 44+0 had broken post but scratch condition cannot be confirmed.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the device 223600002, apg+ sizer pin guide 44+0 would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to end of life problem identified and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot = the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > (b)(4) investigation summary = > according to the information received, "informed via email that the plastic guides are damaged and will need replacing" the product was not returned to depuy synthes, however photos were provided for review.See attachment "(b)(4)" the photo investigation revealed that the device (b)(4), apg+ sizer pin guide 44+0 was worn out, the hex components of the guide have been burred out so that they no longer fit onto the hex component of the guide handle but scratch condition cannot be confirmed.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the device (b)(4), apg+ sizer pin guide 44+0 would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to end of life problem identified and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot = > the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|
|