| Model Number |
FREDX5015-PMA |
| Medical Device Problem Codes |
Retraction Problem (1536); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported through a fred x clinical study, the fred x deployed as intended at the distal landing zone, opened across the neck, but failed to open along the proximal curvature.The fred x could not be captured, retracted within the microcatheter.The device was successfully recaptured through the advancement of the intermediate catheter.A pipeline stent was placed successfully.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned to the manufacturer for analysis; therefore, the alleged product issue cannot be verified.A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.If additional information is received a supplemental mdr report will be provided.The instructions for use (ifu) identifies device retraction difficulty as potential complications associated with use of the device.
|
| |
|
Search Alerts/Recalls
|