| Catalog Number |
51-100050 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported by the distributorship that while inspecting their stock, the sterile packaging was found to be damaged.Attempts have been made and no further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-03049; 0001825034-2023-03050; 0001825034-2023-03051; 0001825034-2023-03052; 0001825034-2023-03053.G2: japan.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility has not been compromised.The reported event has been confirmed by evaluation of the returned product and provided photos.Dhr was reviewed and no discrepancies related to the reported event were found.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional event information to report at this time.
|
| |
|