| Catalog Number |
1400250-15 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1 facility name - (b)(6) hospital.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 90% stenosed lesion in the proximal left circumflex (plcx) artery with moderate calcification and heavy tortuosity.The 2.5x15mm nc trek neo balloon dilatation catheter (bdc) was advanced to the target lesion with resistance due to the anatomy.The balloon ruptured during the first inflation at 12 atmospheres after 5 seconds; therefore, the bdc was removed from the patient.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|
| |
|