| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® no additive (z) tubes have a poor vacuum and it is difficult to puncture the rubber stopper with the needle.There was no report of impact to patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, 100 retention samples from bd inventory were evaluated by visual examination and 10 retention samples by functional testing.No issues were observed relating to the failure modes underfill or difficult to pierce stopper as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure modes underfill and difficult to pierce stopper.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported when using the bd vacutainer® no additive (z) tubes have a poor vacuum and it is difficult to puncture the rubber stopper with the needle.There was no report of impact to patient or user.
|
| |
|