Catalog Number LT200
Medical Device Problem Codes Unintended Ejection (1234); Difficult to Insert (1316); Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a mastectomy procedure, when remove the clip bank from the sterile packaging, it already falls apart.The clips can then no longer be properly loaded.No patient harm nor significant delay reported.Procedure finished with same like device.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 12/22/2023 d4: batch # unk investigation summary the product was returned to ethicon endo surgery for evaluation.Visual inspection was conducted on the returned reload.Visual analysis of the returned sample determined that the lt200 (b) cartridge was received empty along with 4 loose clips; 3 of them were malformed in u,s, and j form and one unformed.No functional test was performed due to condition of the returned clips.The event reported was confirmed and it is related to improper use of the device.Select the appropriate size ligaclip extra ligating clip cartridge and corresponding clip applier.With the cartridge slots facing away from the base, insert the cartridge into one of the channel openings of the lc800 stainless steel cartridge base.Ensure that the cartridge is past the channel opening and securely held in place.Grasp the applier in the box lock area using pencil grip technique.If using an endoscopic applier, grasp the applier by the handle and trigger.Insert the jaws of the instrument into the individual cartridge slot, making sure the tips of the applier are perpendicular to the surface of the cartridge.Insert the applier until it stops.Do not force the applier.It should enter and withdraw from the cartridge smoothly.Keeping the applier jaws perpendicular to the surface of the cartridge, retract the applier from the cartridge.The clip will be securely held in the applier jaws.It is not necessary to maintain ring or trigger tension to hold the clip in the applier jaws.Position the clip around the tubular structure to be ligated.Apply sufficient force to fully close the applier to assure that the clip is satisfactorily placed and secured.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key18388591
Report Number3005075853-2023-09760
Device Sequence Number19825370
Product Code FZP
UDI-Device Identifier10705036012801
UDI-Public10705036012801
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K830503
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberLT200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2023
Initial Report FDA Received Date12/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-