BOSTON SCIENTIFIC CORPORATION ENCORE 26; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number 3040
Medical Device Problem Codes Break (1069); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3.Used first day of the aware date as the event date was not reported.
 
Event or Problem Description
It was reported that the gauge was showing incorrect pressure.An encore 26 inflation device was selected for use.The encore device inflated the balloon even though the pressure gauge needle did not move.Additionally, a loose jingling sound was heard when the indeflator was moved which may indicate that there is a broken part within.The procedure was completed with another of the same device.There were no patient complications reported.
 
Event or Problem Description
It was reported that the gauge was showing incorrect pressure.An encore 26 inflation device was selected for use.The encore device inflated the balloon even though the pressure gauge needle did not move.Additionally, a loose jingling sound was heard when the indeflator was moved which may indicate that there is a broken part within.The procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
B3.Used first day of the aware date as the event date was not reported.Device evaluated by manufacturer: the device was returned for analysis.A visual and functional analysis were performed on the device.The gauge needle was at 0 atm when received.The device does not have visual defects.The unit passed all functional tests without issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE 26
Common Device Name
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18389169
Report Number2124215-2023-69753
Device Sequence Number3967468
Product Code MAV
UDI-Device Identifier08714729177029
UDI-Public08714729177029
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K955869
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/27/2025
Device Model Number3040
Device Catalogue Number3040
Device Lot Number0031913176
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/04/2023
Supplement Date Received by Manufacturer01/23/2024
Initial Report FDA Received Date12/22/2023
Supplement Report FDA Received Date02/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-