| Model Number |
3040 |
| Medical Device Problem Codes |
Break (1069); Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3.Used first day of the aware date as the event date was not reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the gauge was showing incorrect pressure.An encore 26 inflation device was selected for use.The encore device inflated the balloon even though the pressure gauge needle did not move.Additionally, a loose jingling sound was heard when the indeflator was moved which may indicate that there is a broken part within.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the gauge was showing incorrect pressure.An encore 26 inflation device was selected for use.The encore device inflated the balloon even though the pressure gauge needle did not move.Additionally, a loose jingling sound was heard when the indeflator was moved which may indicate that there is a broken part within.The procedure was completed with another of the same device.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3.Used first day of the aware date as the event date was not reported.Device evaluated by manufacturer: the device was returned for analysis.A visual and functional analysis were performed on the device.The gauge needle was at 0 atm when received.The device does not have visual defects.The unit passed all functional tests without issues.
|
| |
|