MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3341202 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
10/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 39-year-old caucasian female patient underwent primary breast augmentation with a 390cc mentor memoryshape breast implant on both sides and experienced capsular contracture; baker grade iv on her right side postoperatively; diagnosed after physical exam.The patient has consulted with the healthcare professional.Per information received on december 7, and 8, 2023, mentor became aware the patient also developed a waterfall effect in addition to right side capsular contracture; baker grade iv.Hence, underwent bilateral mastopexy revision with catalog number smhb400; serial number (b)(6), and with catalog number smhb400; serial number (b)(6) on (b)(6) 2023.This medwatch report is for the patient¿s left-sided device.
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: left waterfall effect mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On december 26, 2023, device evaluation was completed.Device evaluation summary: mentor conducted a visual inspection of the returned device.During visual evaluation, no apparent damage or visual anomalies were observed on the mentor memoryshape¿ mh 390cc returned device.Some breast ptosis is a normal component of the mature breast.Ptosis becomes undesirable when the breast parenchyma predominates below the areola, droops considerably below the inframammary fold and the nipple points downward.Augmentation alone, without consideration of the ptosis can produce a less than desirable cosmetic result known as a "rock in sock" deformity or an increased ptotic appearance.Ptosis is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This supplemental medwatch report is for the patient¿s left-sided device.Manufacturer¿s reference number: (b)(4).
|
| |
|