BD MEDICAL - DIABETES CARE BD INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE; PISTON SYRINGE Back to Search Results
Catalog Number 328418
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that while using the bd insulin syringes with bd ultra-fine¿ needle the stopper was stuck inside the barrel.The following information was provided by the initial reporter: consumer reported, the "stopper" is stuck inside the barrel, preventing plunger rod from moving to draw insulin.Stated, cannot use syringe.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval yes.D9: returned to manufacturer on: 05-jan-2024.H.6.Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was not able to duplicate or confirm the indicated issue and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.H3 other text : see h.10.
 
Event or Problem Description
It was reported that while using the bd insulin syringes with bd ultra-fine¿ needle the stopper was stuck inside the barrel.The following information was provided by the initial reporter: consumer reported, the "stopper" is stuck inside the barrel, preventing plunger rod from moving to draw insulin.Stated, cannot use syringe.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key18390573
Report Number1920898-2023-00849
Device Sequence Number8724064
Product Code FMF
UDI-Device Identifier00382908418034
UDI-Public(01)00382908418034
Combination Product (Y/N)Y
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K190054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number328418
Device Lot Number3058666
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer03/15/2024
Initial Report FDA Received Date12/25/2023
Supplement Report FDA Received Date03/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-