• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC NATURA; PROTECTOR, OSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC NATURA; PROTECTOR, OSTOMY Back to Search Results
Model Number 413180
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device 31 of 63.An investigation was completed for this complaint by the assigned manufacturing site investigation team.The complaint summary of this investigation required rework.The complaint required to create an initial 31st emdr out of 63 emdrs that was missed initially.A capa was raised and an updated/corrected investigation summary has been completed in accordance with assigned capa.A supplemental mdr is being submitted to document the completion of investigation rework and final investigation; the complaint record will proceed to closure.E1: complainant information: (b)(6).Based on the available information, this event is deemed to be a reportable malfunction.Batch records review results: batch records of final products and bulking lots were reviewed to confirm if all requirements were followed.Results of the batch records review were documented in each individual section of the 6 m¿s methodology.All impacted lots were documented in the attachment of final products with bulking lots.Sample/photo analysis results: no samples were received for evaluation/one photo was received within complaints reported.Conclusion summary of the related event: based on the investigation findings through revision of the batch records documentation, process observation, personnel interviewed, methodology implemented and defects simulation, the specific root cause could not be identified.Nevertheless, there are seven (07) conditions that could be considered, based on objective evidence and expertise, as contributor factors for this event and, one (01) opportunity for improvement were found: 1.Materials investigations: fabric rolls not rolled uniformly, no welded or over welded flanges.2.Process/methods investigation: wrong adhesive tape used for joint fabrics rolls.Opportunities for improvement, introduce quality control (qc) tooling used for quality inspection purpose in the calibration program to guarantee measurement assurance.3.Machinery / equipment/ software investigation: different movement relation vs cardan in the yamaha arm pads.If the cognex cameras of the vision system is not clean.In the vision system, the model in the system is not centralized.When the yamaha is not adjusted if the web index is moved, due to the positioning of the vision system.No issues were identified for manpower, measurement, and environment investigations actions will be taken for each factor and are going to be summarized on corrective and preventive actions (capa) plan.The investigation has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number.Reporting site: 1049092.Manufacturing site: 9618003.
 
Event or Problem Description
It was reported the ring of pouch is not located in the middle of 63 wafers.Per the photos received, the starter hole is off center.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATURA
Common Device Name
PROTECTOR, OSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR  91000
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal 91000
DR   91000
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 
3365424679
MDR Report Key18391455
Report Number9618003-2023-15375
Device Sequence Number2932481
Product Code EXE
UDI-Device Identifier00768455124503
UDI-Public00768455124503
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date01/02/2023
Device Model Number413180
Device Lot Number8A06572
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2018
Initial Report FDA Received Date12/26/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-