Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code
No Clinical Signs, Symptoms or Conditions (4582)
Date of Event
10/26/2021
Type of Reportable Event
Malfunction
Additional Manufacturer Narrative
As reported by the technical personnel, the root cause, in this case, can be linked to the temperature sensor being disconnected.No device-related death or serious injury was reported.However, if it were to recur, risk of overheating due to a disconected component inside the device could affect the device in a catastrophic manner that may lead to a serious injury.
Event or Problem Description
A customer informed that a new device showed an intense smoke smell.No further information was provided as if it was out of the box or as a result of the firs use.