Model Number M00513840
Medical Device Problem Codes Positioning Failure (1158); Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/15/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block e1: the initial reporter's phone number is (b)(6).Block h6: imdrf device code a15 captures the reportable investigation results of stent partially deployed.Block h10: an ultraflex esophageal proximal covered stent and delivery system were received for analysis.The stent was received partially deployed.Visual inspection found the shaft kinked.During functional inspection, it was not possible to deploy the stent by pulling the finger ring as the shaft bowed due to the kink on the shaft; however, the stent was able to be deployed by pulling the black deployment suture distal to the delivery handle.The reported event of stent failure to deploy was confirmed because the stent was received partially deployed.It is most likely that procedural factors such as lesion characteristics, handling of the device, and the technique used by the physician (force applied), could have contributed to the observed problem of shaft kinked which resulted in the stent being unable to fully deploy during the procedure.Therefore, a review and analysis of all available information indicated the most probable cause is adverse event related to procedure.A product labeling review identified that the device was used per the instructions for use (ifu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that an ultraflex esophageal proximal covered stent was used in the middle esophagus to treat malignant stenosis and dysphagia during a stent placement procedure performed on (b)(6) 2023.The patient's anatomy was tortuous and was dilated prior to stent placement.During the procedure, the stent could not be deployed inside the patient.The stent was fully covered by the deployment suture when it was removed from the patient.The procedure was completed with another ultraflex esophageal stent.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.Note: this event has been deemed an mdr reportable event based on the investigation results which revealed that the ultraflex esophageal stent was partially deployed.Please see block h10 for full investigation details.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX ESOPHAGEAL NG
Common Device Name
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18393339
Report Number3005099803-2023-06864
Device Sequence Number21024494
Product Code ESW
UDI-Device Identifier08714729716099
UDI-Public08714729716099
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K091816
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/08/2024
Device Model NumberM00513840
Device Catalogue Number1384
Device Lot Number0029915970
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/05/2023
Initial Report FDA Received Date12/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient Weight72 KG
-
-