| Model Number |
M00513840 |
| Medical Device Problem Codes |
Positioning Failure (1158); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block e1: the initial reporter's phone number is (b)(6).Block h6: imdrf device code a15 captures the reportable investigation results of stent partially deployed.Block h10: an ultraflex esophageal proximal covered stent and delivery system were received for analysis.The stent was received partially deployed.Visual inspection found the shaft kinked.During functional inspection, it was not possible to deploy the stent by pulling the finger ring as the shaft bowed due to the kink on the shaft; however, the stent was able to be deployed by pulling the black deployment suture distal to the delivery handle.The reported event of stent failure to deploy was confirmed because the stent was received partially deployed.It is most likely that procedural factors such as lesion characteristics, handling of the device, and the technique used by the physician (force applied), could have contributed to the observed problem of shaft kinked which resulted in the stent being unable to fully deploy during the procedure.Therefore, a review and analysis of all available information indicated the most probable cause is adverse event related to procedure.A product labeling review identified that the device was used per the instructions for use (ifu) / product label.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an ultraflex esophageal proximal covered stent was used in the middle esophagus to treat malignant stenosis and dysphagia during a stent placement procedure performed on (b)(6) 2023.The patient's anatomy was tortuous and was dilated prior to stent placement.During the procedure, the stent could not be deployed inside the patient.The stent was fully covered by the deployment suture when it was removed from the patient.The procedure was completed with another ultraflex esophageal stent.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.Note: this event has been deemed an mdr reportable event based on the investigation results which revealed that the ultraflex esophageal stent was partially deployed.Please see block h10 for full investigation details.
|
| |
|
Search Alerts/Recalls
|