| Model Number |
SAW ATTACHMENT-SAGITTAL |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 11/28/2023, it was reported by a sales representative via sems-06411874 that an ar-400sag drillsaw sports 400 system was not moving as rapidly as it should and would not cut.This occurred during a case, with no effect on the patient.There was no additional information provided, and additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint allegation is confirmed.One unpackaged ar-400sag serial/batch number (b)(6) was received for investigation.No functional testing was performed for this device due to no mating part.Visual evaluation revealed a worn-out crew at the retaining claw.Signs of wear and tear, as well as scratch marks on the head and body of the device.The most likely cause of the reported failure is the use error of mishandling the device.
|
| |
|