| Model Number |
ACL TIGHTROPE RT W/ DEPLOYING SUTURE |
| Medical Device Problem Codes |
Positioning Failure (1158); Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 11/28/2023, it was reported by an arthrex subsidiary employee via email that an ar-1588rt-j acl tightrope rt would not tension.After the graft was implanted, the tibial fixation and tensioning were loose and after several pulls, the graft fell off.It was confirmed that the graft did not tear, so another ar-1588rt-j was used to complete fixation.This was discovered during a procedure on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information.Complaint allegation is not confirmed.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
|
| |
|