• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC MOD SQUARE EXP 5.5 SI DRIVER; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC MOD SQUARE EXP 5.5 SI DRIVER; SCREWDRIVER Back to Search Results
Catalog Number 698341353
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.E3: reporter is a j&j sales representative.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the torque shafts x2 & screwdriver shafts x2 all stripped during our revision case.Opened peel packs.Fragments generated.There was no surgical delay.The procedure completed successfully.There was no patient involvement.This report is for one (1)mod square exp 5.5 si driver this is report 2 of 4 for complaint x25 final tightener.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A review of the receiving inspection (ri) for mod square exp 5.5 si driver was conducted identifying that lot number gm56219mt released in one batch.- batch1: lot qty of (b)(4) units were released on 13 apr 2020 with no discrepancies.Supplier: (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The photo investigation revealed that the tip of mod square exp 5.5 si driver was stripped and device shows the signs of usage.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned device does not found signs of stripped at the tip of the mod square exp 5.5 si driver.However the device was broken from the tip.The broken fragment was not returned for evaluation.A dimensional inspection was not performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.The overall complaint was unconfirmed as the observed condition of the mod square exp 5.5 si driver would not contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.¿based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOD SQUARE EXP 5.5 SI DRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key18394353
Report Number1526439-2023-02604
Device Sequence Number16876997
Product Code HXX
UDI-Device Identifier10705034473703
UDI-Public(01)10705034473703
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number698341353
Device Lot NumberGM56219MT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/04/2023
Supplement Date Received by Manufacturer02/08/2024
Initial Report FDA Received Date12/26/2023
Supplement Report FDA Received Date02/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/13/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
X25 FINAL TIGHTENER.; X25 FINAL TIGHTENER.
Patient SexUnknown
-
-