• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYSHAPE BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3541358
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Crushing Injury (1797); Deformity/ Disfigurement (2360)
Date of Event 10/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent primary breast augmentation with two 395cc mentor memoryshape breast implants.Post-operatively, the patient was diagnosed, via mammogram/ultrasound, with left breast implant rupture.The patient also fell prior to the rupture.As a result, the patient underwent breast implant removal surgery on (b)(6) 2023.
 
Additional Manufacturer Narrative
On december 29, 2023, mentor received additional information indicating that the patient was actually diagnosed with bilateral breast implant ruptures during the revision surgery.In addition, the patient's implants were replaced with the following devices: (left) 270cc mentor memorygel xtra breast implant catalog: smpx270 lot: 9955316 sn: (b)(6) and (right) 270cc mentor memorygel xtra breast implant catalog: smpx270 lot: 9964615 sn: (b)(6).On january 3, 2024, the mentor failure analysis lab received the device for evaluation.On january 5, 2024, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the mentor memoryshape¿ mm 395cc breast implant was found to be ruptured.The evaluation determined that the possible cause of the rupture is the trauma experienced.Rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Trauma is a known inherent risk of gel-filled mammary prosthesis.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.No corrective and preventive action (capa) is required now.This medwatch form is for the left breast prosthesis.Rupture on the right breast implant will be reported under mrn: 1645337-2024-00543.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYSHAPE BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key18394743
Report Number1645337-2023-15124
Device Sequence Number12942619
Product Code FTR
UDI-Device Identifier00081317005483
UDI-Public00081317005483
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P060028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2018
Device Catalogue Number3541358
Device Lot Number6757225
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/30/2023
Supplement Date Received by Manufacturer12/29/2023
Initial Report FDA Received Date12/26/2023
Supplement Report FDA Received Date01/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/07/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceAsian
-
-