| Model Number |
3854 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was fractured.The target lesion was located in the mildly tortuous and moderately calcified artery.A 10mmx2.50mm wolverine cutting balloon was selected for use.During the procedure, the balloon was loaded onto a run through and the doctor went to advance through the copilot when the mid shaft got separated.The device was removed from the patient's body successfully and the procedure was completed with a different device.No complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was fractured.The target lesion was located in the mildly tortuous and moderately calcified artery.A 10mmx2.50mm wolverine cutting balloon was selected for use.During the procedure, the balloon was loaded onto a run through and the doctor went to advance through the copilot when the mid shaft got separated.The device was removed from the patient's body successfully and the procedure was completed with a different device.No complications were reported.It was further reported that the device was outside of the body at the time of the event and it never entered the body of the patient as this was noted during preparation.Also, the break was roughly 30-40% up the shaft from the balloon.
|
| |
|