BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3854
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was fractured.The target lesion was located in the mildly tortuous and moderately calcified artery.A 10mmx2.50mm wolverine cutting balloon was selected for use.During the procedure, the balloon was loaded onto a run through and the doctor went to advance through the copilot when the mid shaft got separated.The device was removed from the patient's body successfully and the procedure was completed with a different device.No complications were reported.
 
Event or Problem Description
It was reported that the device was fractured.The target lesion was located in the mildly tortuous and moderately calcified artery.A 10mmx2.50mm wolverine cutting balloon was selected for use.During the procedure, the balloon was loaded onto a run through and the doctor went to advance through the copilot when the mid shaft got separated.The device was removed from the patient's body successfully and the procedure was completed with a different device.No complications were reported.It was further reported that the device was outside of the body at the time of the event and it never entered the body of the patient as this was noted during preparation.Also, the break was roughly 30-40% up the shaft from the balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON
Common Device Name
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18395174
Report Number2124215-2023-73778
Device Sequence Number14186140
Product Code NWX
UDI-Device Identifier08714729888246
UDI-Public08714729888246
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P950020
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/01/2025
Device Model Number3854
Device Catalogue Number3854
Device Lot Number0031156920
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/15/2023
Supplement Date Received by Manufacturer01/04/2024
Initial Report FDA Received Date12/26/2023
Supplement Report FDA Received Date01/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-