Model Number NV GXL LINER LIPPED 36MM ID GROUP 2
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Date of Event 12/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately six years post initial tha, the male patient was revised due to pain.Poly wear was noted by surgeon.A new vit e liner and ceramic head were implanted.There was no issues with surgery.Photos received.The devices are not available for evaluation.
 
Additional Manufacturer Narrative
Section d10: concomitant products.Biolox delta femoral head 36mm od, +0mm (cat# 170-36-00 / serial# (b)(6).Alteon 6.5mm screw, 30mm (cat# 180-65-30 / serial# (b)(6).Integrip cc, cluster 54mm, g2 (cat# 186-01-54 / serial# (b)(6).Wedge plasma x/o sz 10 (cat# 188-01-10 / seral# (b)(6).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
H3.Investigation results- based on the available information, the patient involved meets risk criteria for early prosthesis wear -combination of largest available femoral head and/or thinnest available acetabular liner was used.The most likely underlying cause for the revision due to early prosthesis wear reported is a combination of risk factors.However, this cannot be confirmed from the reported information and the devices were not available for evaluation.These devices are used for treatment not diagnosis.There is no other information available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: e1, e2, e3, g2, h6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LINER LIPPED 36MM ID GROUP 2
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
matt collins
MDR Report Key18395188
Report Number1038671-2023-03055
Device Sequence Number18280047
Product Code JDI
UDI-Device Identifier10885862207036
UDI-Public10885862207036
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K121392
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/04/2023
Device Model NumberNV GXL LINER LIPPED 36MM ID GROUP 2
Device Catalogue Number132-36-52
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2023
Supplement Date Received by Manufacturer04/15/2024
04/24/2026
Initial Report FDA Received Date12/26/2023
Supplement Report FDA Received Date04/18/2024
05/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/06/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1732-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-