• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 3-POINT SHOULDER DISTRACTION SYSTEM; TRACTION APPARATUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 3-POINT SHOULDER DISTRACTION SYSTEM; TRACTION APPARATUS Back to Search Results
Model Number 3-POINT SHOULDER DISTRACTION SYSTEM
Medical Device Problem Codes Defective Device (2588); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/28/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
On 11/28/2023, it was reported by a sales representative via (b)(4) that an ar-1600m 3-point shoulder distraction system looked like the crimp failed, the weights hit the ground and the patient's arm dropped onto the sterile field.This occurred during a case.There was no additional information provided and additional information has been requested.
 
Additional Manufacturer Narrative
Additional information: h6.The complaint was confirmed.One unpackaged ar-1600m 3-point shoulder distraction system batch 15032122 was received for evaluation.Upon assembly and visual inspection of the device, it was noted that the shoulder suspension tower wire crimp was damaged.Also, it was noted that the red protector of the wire was broken.No functional test was performed due to the damage to the device.The reported condition is most likely caused by overstressing a device that is damaged naturally and inevitably as a result of normal wear or aging.
 
Event or Problem Description
Additional information was provided on 12/13/2023 where the sales representative stated that this event occurred during a left shoulder arthroscopy on (b)(6)2023.They used tape to secure the weights.
 
Additional Manufacturer Narrative
Additional information was provided on 12/13/2023 where the sales representative stated that this event occurred during a left shoulder arthroscopy on (b)(6)2023.They used tape to secure the weights.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3-POINT SHOULDER DISTRACTION SYSTEM
Common Device Name
TRACTION APPARATUS
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
emily shafer
8009337001
MDR Report Key18395387
Report Number1220246-2023-09646
Device Sequence Number3968651
Product Code HST
UDI-Device Identifier00888867019911
UDI-Public00888867019911
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3-POINT SHOULDER DISTRACTION SYSTEM
Device Catalogue NumberAR-1600M
Device Lot Number15032122
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2023
Supplement Date Received by Manufacturer11/28/2023
12/13/2023
Initial Report FDA Received Date12/26/2023
Supplement Report FDA Received Date02/26/2024
12/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-