| Model Number |
MU-651RA |
| Medical Device Problem Codes |
No Audible Alarm (1019); Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The biomedical engineer (bme) reported that their bedside monitor (bsm) did not have an audible alarm but had a visual alarm.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The biomedical engineer (bme) reported that their bedside monitor (bsm) did not have an audible alarm but had a visual alarm.The alarm could be seen and heard at the central nurses station (cns) and remote network station (rns).Technical support (ts) instructed them to turn the volume up and down to see if they heard any tones and reboot the bsm, but the issue persisted.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer (bme) reported that their bedside monitor (bsm) did not have an audible alarm but had a visual alarm.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of complaint: the biomedical engineer (bme) reported that their bedside monitor (bsm) did not have an audible alarm but had a visual alarm.The alarm could be seen and heard at the central nurses station (cns) and remote network station (rns).Technical support (ts) instructed them to turn the volume up and down to see if they heard any tones and reboot the bsm, but the issue persisted.No patient harm was reported.Investigation summary: we confirmed the issue was attributed to an issue with the monitor, as the speakers on the monitor were likely damaged, (possibly due to wear and tear damage and/or physical damage).In this case, the device was installed at the facility on (b)(6) 2008 and the warranty expired on 10/22/2013, indicating the device has aged approximately 15 (fifteen years), as such, the device is susceptible to wear and tear damage over time.The monitor was replaced to resolve the issue, and no further issues were reported.We have also identified several potential causes of audio sound issues, which are not limited to that speakers can become blown over time.Audio issues may still occur due to physical damage on the cable due to normal wear and tear or due to human factors.To mitigate such issues, any damaged hardware, monitors (with built in speakers) and other connected speakers, cables or hardware components can be replaced to mitigate damage issues.A review of historical data for the device did not reveal any trends that would contribute to component failure that is related to the design or manufacturing of the device.Moreover, this was found to be the only issue at the facility in the past 3 (three) years for this device and issue.In this case, the issue was resolved through ts and education provided to the customer.There is no evidence of an nk device malfunction that may have contributed to the reported issue.Trending will continue to be monitored.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi related data quality updates.Corrected information: d1 brand name: corrected the brand name from bsm-6501a to life scope tr.D4 additional device information / model #: corrected the model # from bsm-6501a to mu-651ra.D4 additional device information / primary unique device identifier (udi) #: corrected the udi # to include the pi information.This is a correction to the suspect medical device involved in the reported event, specifically the unique device identifier (udi) information in section d of the fda form 3500a.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer (bme) reported that their bedside monitor (bsm) did not have an audible alarm but had a visual alarm.No patient harm was reported.
|
| |
|